Time Electronics 7018

 
The 7018 is a differential pressure calibration system suitable for field and laboratory calibration work of both diffential pressure and electrical process loop signals. The instrument is designed for very stable measurement of differential pressures between its two output ports, enabling the calibration of differential pressure gauges, transducers, and transmitters. There are five available versions, according to the required range, either 0.2, 2, 5, or 10bar.

Features include a 4.5 digit LCD display with 4 selectable pressure units plus mA, zeroing switch, and two quick release connector ports (Hi and Lo for differential pressure input) that are situated on the front panel of the unit. Best accuracy is 0.04% of full scale.

For process signal calibration the loop signal is displayed on the LCD display to 0.05% accuracy. Loop drive supply (24V or 36V) is also provided. Both pressure and loop signal can be displayed at the same time to speed up the calibration of pressure transmitters. Additionally a continuity function is available for testing for open circuit loops.

Power is via internal rechargeable cells (24 hours typical use) with a built-in mains charger. The unit can be powered directly from the mains if required. The 7018 is housed in an impact resistant safety case that provides protection and increased durability. Pressure connections are on the front panel and the instrument is supplied with a pressure hose, electrical test lead, 2 electrical probes and 2 crocodile clips.

Options For the 7018, input pressure can be generated by the 7090 pneumatic hand pump (-950mbar to 40bar). Additional pumps and accompanying gauges are also available.

For process control signal simulation Time Electronics’ handheld 7006 Loop-Mate is available as an option. It provides 4 - 20mA and 0 - 10V signals at levels of 0%, 10%, 25%, 50%, 75%, 90%, 100%. It can also step through these set points automatically to allow hands free calibration. It is battery powered and can be stored in the 7018’s supplied carry pouch.

The 7018 is fitted with RS-232 connection as standard. Display readback can be performed using EasyCal calibration software. With comprehensive features EasyCal can simplify and speed up calibration procedures. Easily produce calibration certificates and reports to international quality standards.
Pressure
Range (bar) 0 to ±0.2 0 to ±2 0 to ±5 0 to ±10
Accuracy (% of fs) 0.1 0.04 0.04 0.04
Resolution (bar) 0.01mb .0001 .001 .001


Units 0.2bar version: mbar, PSI, kg/cm2, cmWg and mA.2, 5, and 10bar versions: bar, PSI, kg/cm2, cmWg and mA.
Maximum Static Pressure 10bar
Sensor Piezo-resistive - stainless steel diaphragm
Over pressure warning 1.2 x range full scale - audio and visual (on LCD) warning
Fittings Quick Release


Electrical
Range Loop current measurement 0 to 200mA
Resolution 10μA
Resistance Loop load 5Ω
Accuracy 0.05% of reading ±1 digit
Loop Drive 24V or 36V switch selectable, 50mA max - isolated and with short circuit protection
Continuity Threshold: 100Ω with audio and visual warning
Terminals 4mm industry standard terminal posts


General Specifications
Power Source Internal rechargeable NiMH batteries, or mains supply (230V or 110V 50/60Hz)
Battery Life 24 hrs of typical usage between charges
Display 7 segment LCD with ‘Low Battery’ warning indicator
Interfaces RS-232 for readback and re-calibration (software not supplied)
Case Structural resin which is weather-proof to IP66 standard
Dimensions W270 x H175 x D250mm
Weight 3kg
Supplied with 1 pressure hose: Quick release fitting. Electrical Test Lead: Twin flexible lead, terminated with 4 off 4mm gold plated plugs. Electrical probes: 2 off low voltage probes and 2 crocodile clips. Mains power lead.
Optional Extras Pneumatic hand pump, gauges, loop simulators, EasyCal calibration software, spare fuses, and calibration certificates traceable to NPL and UKAS
Country of Origin UK