A precision calibrator designed to provide rapid high accuracy calibration of portable appliance testers and insulation/continuity testers.
The 5080 PAT calibrator has functions for earth bond, insulation, leakage, touch leakage, and load test.
Housed in a robust safety case the 5080 is easily portable and ideal for field calibration work. It is battery powered, which ensures isolation and prevents inaccuracies due to stray leakage. Features include safety interlock that prevents contact with earth bond studs during insulation and leakage tests, and an LCD display for voltage and current.
The 5080 is a dedicated solution for PAT testers and part of the electrical testing calibration range from Time Electronics. Other models include the 5030 electrical tester calibrator for RCD, loop, insulation, earth testers and multifunction installation testers. For insulation testers the 5069 InsCal is a portable instrument with resistance up to 100GOhm, and 10kV operation.
Technical Specifications
Earth Bond: Calibrated Low Value Resistive Load
Values / Accuracy |
20mΩ, accuracy 5%190mΩ, 210mΩ, 950mΩ, 1.050mΩ, 1.08Ω, 2Ω, 10Ω, 18Ω, accuracy 1% |
Open circuit voltage measurement |
Range: 0 to 20VAccuracy: 0.25% of full scale |
High current AC measurement |
Range: 0 to 50A ACAccuracy: 0.5% of reading +/- 0.2A |
Low current AC measurement |
Range: 0 to 2A ACAccuracy: 0.5% of reading +/- 2mA |
Low current DC measurement |
Range: 0 to 2A DCAccuracy: 1% of reading +/- 2mA |
Insulation: Calibrated High Value Resistive Load
alues / Accuracy |
95kΩ, 105kΩ, 500kΩ, 950kΩ, 1.05MΩ, 5MΩ, 10MΩ, accuracy 0.1%20MΩ, 50MΩ, 100MΩ, accuracy 1.0% |
Open circuit voltage measurement |
Range: 0 to 1kV DCAccuracy: 0.25% of full scale |
Current measurement |
Range: 0 to 20mA DC. Accuracy: 0.25% of full scale |
Leakage: Resistive Load for Leakage Currend
Nominal Leakage Currents |
At 230V AC of 0.5, 1, 2, 5, 10, 15mA |
Mains voltage measurement |
Range: 0 to 500V ACAccuracy 0.25% of full scale |
Leakage current measurement |
Range: 0 to 20mA ACAccuracy 0.25% of full scale |
Touch Leakage: Constant Current Source for Touch Leakage Current
urrent |
1.9mA AC 50 +/-5 Hz, into 2K2 maximum load resistance |
Current measurement |
Range: 0 to 2mA AC. Accuracy: 0.25% of full scale |
Load Test: Connection of External Load VIA AN IEC Connector
Mains voltage measurement |
Range: 0 to 500V ACAccuracy: 0.25% of full scale |
Load current measurement |
Range: 0 to 13A ACAccuracy: 0.5% of full scale |
General Specifications
Power |
Internal battery 6V re-chargeable NiCad, >200 hrs between charges. |
Dimensions |
W406 x H175 x D330mm |
Weight |
5.5kg |
Supplied With |
4mm terminal box adaptor, 110V adaptor, IEC connector (for external load), battery charger, and user manual. |
Optional Extras |
Calibration certificates traceable NPL and UKAS |
Country of Origin |
UK |