Cropico DO5002
The DO5002 micro ohmmeter is a truly remarkable instrument and ideally suited for laboratory testing of samples, for production line testing, and with the addition of the interfaces RS232 / IEEE-488 or PLC, the DO5002 micro ohmmeter may be remotely controlled and integrated into an automated testing system.
The DO5002 micro ohmmeter encompasses all the features required of an ohmmeter in one instrument. Programmable measuring current in 100 steps ensures that the DO5002 is suitable for all your applications. Automatic temperature compensation and Hi / Lo limits permits sorting of components with the minimum of fuss. Statistical analysis of these values allows you to display the max/min /average values as well as peak to peak and standard deviation. The DO5002 will even calculate cable lengths in ohms/metre.
The DO5002 micro ohmmeter encompasses all the features required of an ohmmeter in one instrument. Programmable measuring current in 100 steps ensures that the DO5002 is suitable for all your applications. Automatic temperature compensation and Hi / Lo limits permits sorting of components with the minimum of fuss. Statistical analysis of these values allows you to display the max/min /average values as well as peak to peak and standard deviation. The DO5002 will even calculate cable lengths in ohms/metre.
Digital microhmmeter with temperature compensation
- High accuracy 0.03%, 6 ranges, 200milliohms to 30kilohms
- Adjustable current ranges,10uA-100mA
- Fast sampling (0.02Hz) at 10u ohm resolution
Technical specification
Resistance ranges
Measurement | 4 wire eliminates lead resistance |
---|---|
Display | LCD Graphics panel with backlight 30,000 count |
Ranges | Manual / Automatic |
Auto Zero | Value Average Hold |
Systems Interface | RS232 fitted as standard, IEEE-488, PLC option |
Terminals | 4mm safety sockets |
Working Temperature | 0°C ... 45 °C |
Storage Temperature | -10°C...+60°C |
Mains Supply | 230 V ±10%, 115 V ±10% 47...63Hz |
Safety | IEC 1010 Class 1 |
Dimensions | 339 x 324 x 131mm (W D H) |
Weight | 8.5kg |
Accessories | Supplied with RS232 interface, mains cord, handbook and certificate of conformity |
Resistance ranges
Range | Resolution | Min. Current | Max. Current | Accuracy at full rated current |
---|---|---|---|---|
30kOhm | 1Ohm | 10mA | 100mA | ±(0.03% Rdg +0.02% FS) |
3kOhm | 100mOhm | 100 mA | 1mA | ±(0.03% Rdg +0.01% FS) |
300Ohm | 10mOhm | 1mA | 10mA | ±(0.03% Rdg +0.01% FS) |
30Ohm | 1mOhm | 10mA | 100mA | ±(0.03% Rdg +0.01% FS) |