CALYS 150

 

CALYS 150 field documenting multifunction calibrator is the top instrument of the range. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument.

CALYS 150 does not only work as a simulator (IN / OUT) but also as a dual channel thermometer (IN / IN) to perform comparison calibration. It calibrates HART transmitters (HART communicator integrated into ACL500 modem) and thermistors.

Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…) and audit trails, CALYS 150 complies with both 21 CFR Part 11 and NADCAP Heat Treatment standards and makes advanced data exploitation and full data traceability easier.

High performances for CALYS 150, for advanced use:

  • Temperature Up to 0.005 % RDG
  • Resistance Up to 0.006 % RDG and 50 KΩ range
  • Current: Up to 0.007 % RDG and 100 mA range + Loop Supply 24 V
  • Voltage: Up to 0.005 % RDG and 50 V range
  • Frequency: Up to 0.01 % RDG and 100 KHz range
  • Pressure: With an external pressure module (comparison calibration with a pressure pump)

Using this user-friendly instrument, calibration tasks can be quickly carried out over the whole process chain. Take the 900 g documenting process calibrator to the field with you during the whole week with 10 calibration procedures stored in the device. Run the procedure after connecting the probes to the instrument (Easy connect system®) and save the results for onsite easy and quick calibration. Back to the office, you can then upload the data on a computer in order to issue customized calibration certificates with dedicated calibration software DATACAL.

IP 54, fully protected by an antichoc rubber holster, CALYS 150 integrates "easyconnect" terminals and a wide backlite display that makes it easy to use in any severe or dark conditions. When used with an external pressure module (ref. ACL433),CALYS 150 can measure and simulate pressure (comparison calibration with a pressure pump).

CALYS 150 has also the capability to drive baths and dry-blocks when associated with the specific cable (ref. ACL600).